Proline Cubemass 300 Modbus RS485
Description of device parameters
Hauser
57
Pulse output 1 to n
Navigation
Expert → Sensor → Measured val. → Output values → PFS output 1 to n → Pulse
output 1 to n (0456–1 to n)
Prerequisite
In the
Operating mode
Pulse
option is selected.
Description
Displays the pulse frequency currently output.
User interface
Positive floating-point number
Additional information
Description
• The pulse output is an open collector output.
• This is configured at the factory in such a way that the transistor is conductive for the
duration of the pulse (NO contact) and is safety-oriented.
NC
NO
t
t
0
1
0
1
A0028726
0
Non-conductive
1
Conductive
NC NC contact (normally closed)
NO NO contact (normally open)
The output behavior can be reversed via the
Invert outp.sig.
the transistor does not conduct for the duration of the pulse.
In addition, the behavior of the output in the event of a device alarm (
Failure mode
parameter (→ 131)) can be configured.
Switch status 1 to n
Navigation
Expert → Sensor → Measured val. → Output values → PFS output 1 to n → Switch
status 1 to n (0461–1 to n)
Prerequisite
The
Switch
option is selected in the
Operating mode
Description
Displays the current switch status of the status output.
User interface
• Open
• Closed
Содержание Proline Cubemass 300