Proline Cubemass 300 Modbus RS485
Description of device parameters
Hauser
131
Additional information
Selection
• Forward flow
Positive flow is output, negative flow is not output.
• Forward/Reverse
Positive and negative flow are output (absolute value), but a distinction is not made
between positive and negative flow.
• Reverse flow
Negative flow is output, positive flow is not output.
• Rev. flow comp.
The flow components outside the span are buffered, balanced and output after a
maximum delay of 60 s.
For a detailed description of the options available, see the
Measuring mode
Examples
For a detailed description of the configuration examples, see the
Measuring mode
Failure mode
Navigation
Expert → Output → PFS output 1 to n → Failure mode (0480–1 to n)
Prerequisite
In the
Operating mode
Pulse
option is selected, and one of
the following options is selected in the
Assign pulse
• Mass flow
• Volume flow
• Correct.vol.flow
• Target mass flow
• Carrier mass fl.
Description
Use this function to select the failure mode of the pulse output in the event of a device
alarm.
Selection
• Actual value
• No pulses
Factory setting
No pulses
Additional information
Description
The dictates of safety render it advisable to ensure that the pulse output shows a
predefined behavior in the event of a device alarm.
Selection
• Actual value
In the event of a device alarm, the pulse output continues on the basis of the current
flow measurement. The fault is ignored.
• No pulses
In the event of a device alarm, the pulse output is "switched off".
NOTICE!
A device alarm is a measuring device error that must be taken seriously. It can
affect the measurement quality such that the quality can no longer be guaranteed. The
*
Visibility depends on order options or device settings