Description of device parameters
Proline Cubemass 300 Modbus RS485
146
Hauser
Factory setting
Mass flow
Assign status
Navigation
Expert → Output → PFS output 1 to n → Assign status (0485–1 to n)
Prerequisite
• The
Switch
option is selected in the
Operating mode
parameter (→ 126).
• The
Status
option is selected in the
Switch out funct
parameter (→ 141).
Description
Use this function to select a device status for the switch output.
Selection
• Partial pipe det
• Low flow cut off
Factory setting
Partial pipe det
Additional information
Options
If empty pipe detection or low flow cut off are enabled, the output is conductive.
Otherwise, the switch output is non-conductive.
Switch-on delay
Navigation
Expert → Output → PFS output 1 to n → Switch-on delay (0467–1 to n)
Prerequisite
• The
Switch
option is selected in the
Operating mode
parameter (→ 126).
• The
Limit
option is selected in the
Switch out funct
Description
Use this function to enter a delay time for switching on the switch output.
User entry
0.0 to 100.0 s
Factory setting
0.0 s
Switch-off delay
Navigation
Expert → Output → PFS output 1 to n → Switch-off delay (0465–1 to n)
Prerequisite
• The
Switch
option is selected in the
Operating mode
parameter (→ 126).
• The
Limit
option is selected in the
Switch out funct
Description
Use this function to enter a delay time for switching off the switch output.
User entry
0.0 to 100.0 s
Factory setting
0.0 s