Description of device parameters
Proline Cubemass 300 Modbus RS485
96
Hauser
User entry
Signed floating-point number
Factory setting
0 kg/s
Additional information
Description
Corrected value = (factor × value) + offset
Mass flow factor
Navigation
Expert → Sensor → Sensor adjustm. → Variable adjust → Mass flow factor (1832)
Description
Use this function to enter a quantity factor (without time) for the mass flow. This
multiplication factor is applied over the mass flow range.
User entry
Positive floating-point number
Factory setting
1
Additional information
Description
Corrected value = (factor × value) + offset
Vol. flow offset
Navigation
Expert → Sensor → Sensor adjustm. → Variable adjust → Vol. flow offset (1841)
Description
Use this function to enter the zero point shift for the volume flow trim. The volume flow
unit on which the shift is based is m³/s.
User entry
Signed floating-point number
Factory setting
0 m³/s
Additional information
Description
Corrected value = (factor × value) + offset
Vol. flow factor
Navigation
Expert → Sensor → Sensor adjustm. → Variable adjust → Vol. flow factor (1846)
Description
Use this function to enter a quantity factor (without time) for the volume flow. This
multiplication factor is applied over the volume flow range.
User entry
Positive floating-point number