Proline Cubemass 300 Modbus RS485
Description of device parameters
Hauser
91
Factory setting
Calc ref density
Additional information
Selection
The
Ref. dens API 53
option is suitable only for applications involving LPG
, where the
flow rate is measured on the basis of the corrected volume flow.
Selecting this option means that the reference density is used, taking into account the
values in table 53 E of API MPMS section 11.2. Temperature measurement (measured
internally or read into the device from an external source→ 87→ 87) and density
measurement take place during operation while the medium is flowing. The mass flow is
divided by the reference density to give the corrected volume flow and is issued as an
output signal.
Ext. ref.density
Navigation
Expert → Sensor → Calculated value → Corr. vol.flow. → Ext. ref.density (6198)
Prerequisite
In the
Corr. vol.flow.
parameter (→ 90), the
Ext. ref.density
option is selected.
Description
Displays the reference density which is read in externally, e.g. via the current input.
User interface
Floating point number with sign
Additional information
Dependency
The unit is taken from the
Ref. dens. unit
Fix ref.density
Navigation
Expert → Sensor → Calculated value → Corr. vol.flow. → Fix ref.density (1814)
Prerequisite
The
Fix ref.density
option is selected in the
Corr. vol.flow.
parameter.
Description
Use this function to enter a fixed value for the reference density.
User entry
Positive floating-point number
Factory setting
1 kg/Nl
Additional information
Dependency
The unit is taken from the
Ref. dens. unit
5)
liquefied petroleum gas