Description of device parameters
Proline Cubemass 300 Modbus RS485
154
Hauser
Additional information
Selection
• Actual status
In the event of a device alarm, faults are ignored and the current behavior of the input
value is output by the relay output. The
Actual status
option behaves in the same way as
the current input value.
• Open
In the event of a device alarm, the relay output's transistor is set to
non-conductive
.
• Closed
In the event of a device alarm, the relay output's transistor is set to
conductive
.
Switch status
Navigation
Expert → Output → Relay output 1 to n → Switch status (0801–1 to n)
Description
Displays the current status of the relay output.
User interface
• Open
• Closed
Additional information
User interface
• Open
The relay output is not conductive.
• Closed
The relay output is conductive.
Powerless relay
Navigation
Expert → Output → Relay output 1 to n → Powerless relay (0816–1 to n)
Description
Use this function to select the quiescent state for the relay output.
Selection
• Open
• Closed
Factory setting
Open
Additional information
Selection
• Open
The relay output is not conductive.
• Closed
The relay output is conductive.