Description of device parameters
Proline Cubemass 300 Modbus RS485
138
Hauser
Description
Use this function to enter a time constant for the reaction time of the output signal to
fluctuations in the measured value.
User entry
0 to 999.9 s
Factory setting
0.0 s
Additional information
User entry
Use this function to enter a time constant (PT1 element
) for frequency output damping:
• If a low time constant is entered, the current output reacts particularly quickly to
fluctuating measured variables.
• On the other hand, the current output reacts more slowly if a high time constant is
entered.
Damping is switched off if
0
is entered (factory setting).
The frequency output is subject to separate damping that is independent of all preceding
time constants.
Response time
Navigation
Expert → Output → PFS output 1 to n → Response time (0491–1 to n)
Prerequisite
In the
Operating mode
Frequency
option is selected and in the
Assign freq.
parameter (→ 132) one of the following options is selected:
• Mass flow
• Volume flow
• Correct.vol.flow
• Target mass flow
• Carrier mass fl.
• Density
• Ref.density
• Concentration
• Temperature
• Carr. pipe temp.
• Electronic temp.
• Osc. freq. 0
• Freq. fluct. 0
• Osc. ampl. 0
• Osc. damping 0
• Osc.damp.fluct 0
• Signal asymmetry
• Exc. current 0
Detailed description of the options
Oscil. frequency, Oscil. amplitude, Oscil.
damping
and
Signal asymmetry
:
Value 1 display
Description
Displays the response time. This specifies how quickly the pulse/frequency/switch output
reaches the measured value change of 63 % of 100 % of the measured value change.
User interface
Positive floating-point number
7)
proportional transmission behavior with first order delay
*
Visibility depends on order options or device settings