Proline Cubemass 300 Modbus RS485
Description of device parameters
Hauser
141
Output freq. 1 to n
Navigation
Expert → Output → PFS output 1 to n → Output freq. 1 to n (0471–1 to n)
Prerequisite
In the
Operating mode
Frequency
option is selected.
Description
Displays the actual value of the output frequency which is currently measured.
User interface
0.0 to 12
500.0 Hz
Switch out funct
Navigation
Expert → Output → PFS output 1 to n → Switch out funct (0481–1 to n)
Prerequisite
The
Switch
option is selected in the
Operating mode
Description
Use this function to select a function for the switch output.
Selection
• Off
• On
• Diag. behavior
• Limit
• Fl. direct.check
• Status
Factory setting
Off
Additional information
Selection
• Off
The switch output is permanently switched off (open, non-conductive).
• On
The switch output is permanently switched on (closed, conductive).
• Diag. behavior
Indicates if the diagnostic event is present or not. Is used to output diagnostic
information and to react to it appropriately at the system level.
• Limit
Indicates if a specified limit value has been reached for the process variable. Is used to
output diagnostic information relating to the process and to react to it appropriately at
the system level.
• Fl. direct.check
Indicates the flow direction (forward or reverse flow).
• Status
Indicates the device status depending on whether empty pipe detection or low flow cut
off is selected.