Proline Cubemass 300 Modbus RS485
Description of device parameters
Hauser
139
Additional information
Description
The response time is made up of the time specified for the following dampings:
• Damping of pulse/frequency/switch output → 120
and
• Depending on the measured variable assigned to the output.
– Flow damping
or
– Density damping
or
– Temperature damping
Failure mode
Navigation
Expert → Output → PFS output 1 to n → Failure mode (0451–1 to n)
Prerequisite
In the
Operating mode
Frequency
option is selected, and one
of the following options is selected in the
Assign freq.
parameter (→ 132):
• Mass flow
• Volume flow
• Correct.vol.flow
• Target mass flow
• Carrier mass fl.
• Density
• Ref.density
• Concentration
• Temperature
• Carr. pipe temp.
• Electronic temp.
• Osc. freq. 0
• Freq. fluct. 0
• Osc. ampl. 0
• Osc. damping 0
• Osc.damp.fluct 0
• Signal asymmetry
• Exc. current 0
Detailed description of the options
Oscil. frequency, Oscil. amplitude, Oscil.
damping
and
Signal asymmetry
:
Value 1 display
Description
Use this function to select the failure mode of the frequency output in the event of a device
alarm.
Selection
• Actual value
• Defined value
• 0 Hz
Factory setting
0 Hz
*
Visibility depends on order options or device settings