Description of device parameters
Proline Cubemass 300 Modbus RS485
100
Hauser
Nominal diameter (2807)
C0 to 5 (6022)
Cal. factor
Navigation
Expert → Sensor → Calibration → Cal. factor (6025)
Description
Displays the current calibration factor for the sensor.
User interface
Signed floating-point number
Factory setting
Depends on nominal diameter and calibration.
Zero point
Navigation
Expert → Sensor → Calibration → Zero point (6195)
Description
Use this function to enter the zero point correction value for the sensor.
User entry
Signed floating-point number
Factory setting
Depends on nominal diameter and calibration.
Nominal diameter
Navigation
Expert → Sensor → Calibration → Nominal diameter (2807)
Description
Displays the nominal diameter of the sensor.
User interface
DNxx / x"
Factory setting
Depends on the size of the sensor
Additional information
Description
The value is also specified on the sensor nameplate.