Description of device parameters
Proline Cubemass 300 Modbus RS485
226
Hauser
Additional information
Description
The desired simulation value is defined in the
Switch status 1 to n
parameter.
Selection
• Off
Relay simulation is switched off. The device is in normal measuring mode or another
process variable is being simulated.
• On
Relay simulation is active.
Switch status 1 to n
Navigation
Expert → Diagnostics → Simulation → Switch status 1 to n (0803–1 to n)
Prerequisite
The
On
option is selected in the
Switch sim. 1 to n
parameter parameter.
Description
Use this function to select a relay value for the simulation. In this way, users can verify the
correct adjustment of the relay output and the correct function of downstream switching
units.
Selection
• Open
• Closed
Additional information
Selection
• Open
Relay simulation is switched off. The device is in normal measuring mode or another
process variable is being simulated.
• Closed
Relay simulation is active.
Puls.outp.sim.
Navigation
Expert → Diagnostics → Simulation → Puls.outp.sim. (0988)
Description
Use this function to switch simulation of the double pulse output on and off. The display
alternates between the measured value and a diagnostic message of the "Function check"
category (C) while simulation is in progress.
Selection
• Off
• Fixed value
• Down-count. val.
Factory setting
Off
Содержание Proline Cubemass 300
Страница 2: ......
Страница 294: ...www addresses endress com...