Proline Cubemass 300 Modbus RS485
Modbus RS485 Register Information
Hauser
269
Navigation: Expert → Output → PFS output 1 to n
Parameter
Register
Data type
Access
Selection / User entry / User interface
→
Assign freq. (0478–1 to n)
1: 2614
2: 2615
3: 9915
Integer
Read / Write
0 = Off
1 = Mass flow
2 = Volume flow
3 = Correct.vol.flow
4 = Density
5 = Ref.density
7 = Temperature
8 = Osc. ampl. 0
9 = Signal asymmetry
23 = Carr. pipe temp.
32 = Exc. current 0
39 = Electronic temp.
48 = Osc. freq. 0
63 = Osc. damping 0
66 = Pressure
67 = Osc.damp.fluct 0
68 = Freq. fluct. 0
73 = Concentration
74 = Target mass flow
75 = Carrier mass fl.
78 = Carrier vol. fl.
79 = Carr.corr.vol.fl
81 = HBSI
83 = Oil density
84 = Water density
86 = GSV flow
87 = GSVa
88 = Oil mass flow
89 = Water mass flow
90 = NSV flow
91 = NSVa
92 = S&W volume flow
93 = Ref.dens.altern.
94 = Oil corr.vol.fl.
95 = Water corr.v.fl.
96 = Targ.corr.vol.fl
97 = Target vol. flow
99 = Oil volume flow
101 = Water cut
102 = Water vol. flow
Min. freq. value (0453–1 to n)
1: 3526 to 3527
2: 3528 to 3529
3: 5767 to 5768
Float
Read / Write
0.0 to 10
000.0 Hz
Max. freq. value (0454–1 to n)
1: 2996 to 2997
2: 2998 to 2999
3: 4710 to 4711
Float
Read / Write
0.0 to 10
000.0 Hz
Val. at min.freq (0476–1 to n)
1: 5887 to 5888
2: 5889 to 5890
3: 5891 to 5892
Float
Read / Write
Signed floating-point number
Val. at max.freq (0475–1 to n)
1: 3514 to 3515
2: 3516 to 3517
3: 5759 to 5760
Float
Read / Write
Signed floating-point number
Measuring mode (0479–1 to n)
1: 2922
2: 2923
3: 4706
Integer
Read / Write
0 = Forward flow
2 = Rev. flow comp.
13 = Forward/Reverse
Damping out. 1 to n (0477–1 to n)
1: 3522 to 3523
2: 3524 to 3525
3: 5763 to 5764
Float
Read / Write
0 to 999.9 s
Response time (0491–1 to n)
1: 5875 to 5876
2: 5877 to 5878
3: 5879 to 5880
Float
Read
Positive floating-point number
Содержание Proline Cubemass 300