Description of device parameters
Proline Cubemass 300 Modbus RS485
84
Hauser
Description
Use this function to enter an upper limit value to enable detection of empty or partially
filled measuring tubes. If the measured density exceeds this value, detection is enabled.
User entry
Signed floating-point number
Factory setting
6
000
Additional information
User entry
The upper limit value must be greater than the lower limit value defined in the
Low value
parameter (→ 83).
The unit depends on the process variable selected in the
Assign variable
parameter
Limit value
If the displayed value is outside the limit value, the measuring device displays the
diagnostic message
S862 Partly filled
.
Response time
Navigation
Expert → Sensor → Process param. → Partial pipe det → Response time (1859)
Prerequisite
One of the following options is selected in the
Assign variable
• Density
• Ref.density
Description
Use this function to enter the minimum length of time (debouncing time) the signal must
be present for the diagnostic message
S862 Partly filled
to be triggered if the
measuring pipe is empty or partially full.
User entry
0 to 100 s
Factory setting
1 s
Max. damping
Navigation
Expert → Sensor → Process param. → Partial pipe det → Max. damping (6040)
Description
Use this function to enter a damping value to enable detection of empty or partially filled
measuring tubes.
User entry
Positive floating-point number
Factory setting
0
Additional information
Description
If oscillation damping exceeds the specified value, the measuring device presumes that the
pipe is partially filled and the flow signal is set to
0