Proline Cubemass 300 Modbus RS485
Description of device parameters
Hauser
227
Additional information
Description
The desired simulation value is defined in the
Pulse value
Selection
• Off
Simulation of the double pulse output is switched off. The device is in normal measuring
mode or another process variable is being simulated.
• Fixed value
Pulses are continuously output with the pulse width specified in the
Pulse width
• Down-count. val.
The pulses specified in the
Pulse value
parameter (→ 227) are output.
Pulse value
Navigation
Expert → Diagnostics → Simulation → Pulse value (0989)
Prerequisite
In the
Puls.outp.sim.
Down-count. val.
option is selected.
Description
Use this function to enter a pulse value for simulation of the double pulse output. In this
way, users can verify the correct adjustment of the double pulse output and the correct
function of downstream switching units.
User entry
0 to 65
535
Dev. alarm sim.
Navigation
Expert → Diagnostics → Simulation → Dev. alarm sim. (0654)
Description
Use this function to switch the device alarm on and off.
Selection
• Off
• On
Factory setting
Off
Additional information
Description
The display alternates between the measured value and a diagnostic message of the
"Function check" category (C) while simulation is in progress.