Proline Cubemass 300 Modbus RS485
Description of device parameters
Hauser
53
Additional information
Description
As it is only possible to display a maximum of 7 digits in the operating tool, the current
counter value is the sum of the totalizer value and the overflow value from the
Tot.
overflow 1 to n
parameter if the display range is exceeded.
In the event of an error, the totalizer adopts the mode defined in the
Failure mode
User interface
The value of the process variable totalized since measuring began can be positive or
negative. This depends on the settings in the
Operation mode
parameter (→ 180).
The unit of the selected process variable is specified for the totalizer in the
Unit
totalizer
parameter (→ 178).
Example
Calculation of the current totalizer reading when the value exceeds the 7-digit display
range of the operating tool:
• Value in the
Totalizer val. 1
parameter: 1
968
457 m
3
• Value in the
Tot. overflow 1
parameter: 1 ⋅ 10
7
(1 overflow) = 10
000
000 [m
3
]
• Current totalizer reading: 11
968
457 m
3
Tot. overflow 1 to n
Navigation
Expert → Sensor → Measured val. → Totalizer → Tot. overflow 1 to n (0910–1 to n)
Prerequisite
A process variable is selected in the
Assign variable
Totalizer 1 to n
submenu.
Description
Displays the current totalizer overflow.
User interface
Integer with sign
Additional information
Description
If the current totalizer reading exceeds 7 digits, which is the maximum value range that
can be displayed by the operating tool, the value above this range is output as an overflow.
The current totalizer value is therefore the sum of the overflow value and the totalizer
value from the
Totalizer val. 1 to n
parameter.
User interface
The unit of the selected process variable is specified for the totalizer in the
Unit
totalizer
parameter (→ 178).
Example
Calculation of the current totalizer reading when the value exceeds the 7-digit display
range of the operating tool:
• Value in the
Totalizer val. 1
parameter: 1
968
457 m
3
• Value in the
Tot. overflow 1
parameter: 2 ⋅ 10
7
(2 overflows) = 20
000
000 [m
3
]
• Current totalizer reading: 21
968
457 m
3