Proline Cubemass 300 Modbus RS485
Description of device parameters
Hauser
95
Progress
Navigation
Expert → Sensor → Sensor adjustm. → Zero point adj. → Progress (2808)
Description
The progress of the process is indicated.
User interface
0 to 100 %
"Variable adjust" submenu
Navigation
Expert → Sensor → Sensor adjustm. → Variable adjust
‣
Variable adjust
Mass flow offset (1831)
Mass flow factor (1832)
Vol. flow offset (1841)
Vol. flow factor (1846)
Density offset (1848)
Density factor (1849)
Corr. vol offset (1866)
Corr. vol factor (1867)
Ref.dens. offset (1868)
Ref.dens. factor (1869)
Temp. offset (1870)
Temp. factor (1871)
Mass flow offset
Navigation
Expert → Sensor → Sensor adjustm. → Variable adjust → Mass flow offset (1831)
Description
Use this function to enter the zero point shift for the mass flow trim. The mass flow unit
on which the shift is based is kg/s.