Description of device parameters
Proline Cubemass 300 Modbus RS485
222
Hauser
Curr.out. 1 to n sim.
Navigation
Expert → Diagnostics → Simulation → Curr.out. 1 to n sim. (0354–1 to n)
Description
Use this function to switch simulation of the current output on and off. The display
alternates between the measured value and a diagnostic message of the "Function check"
category (C) while simulation is in progress.
Selection
• Off
• On
Factory setting
Off
Additional information
Description
The desired simulation value is defined in the
Value curr.out 1 to n
parameter.
Selection
• Off
Current simulation is switched off. The device is in normal measuring mode or another
process variable is being simulated.
• On
Current simulation is active.
Value curr.out 1 to n
Navigation
Expert → Diagnostics → Simulation → Value curr.out 1 to n (0355–1 to n)
Prerequisite
In the
Curr.out. 1 to n sim.
parameter, the
On
option is selected.
Description
Use this function to enter a current value for the simulation. In this way, users can verify
the correct adjustment of the current output and the correct function of downstream
switching units.
User entry
3.59 to 22.5 mA
Additional information
Dependency
The input range is dependent on the option selected in the
Current span
parameter
FreqOutputSim 1 to n
Navigation
Expert → Diagnostics → Simulation → FreqOutputSim 1 to n (0472–1 to n)
Prerequisite
In the
Operating mode
Frequency
option is selected.