Proline Cubemass 300 Modbus RS485
Description of device parameters
Hauser
151
• NSVa
• S&W volume flow
• Water cut
• Oil density
• Water density
• Oil mass flow
• Water mass flow
• Oil volume flow
• Water vol. flow
• Oil corr.vol.fl.
• Water corr.v.fl.
• Concentration
• Temperature
• Totalizer 1
• Totalizer 2
• Totalizer 3
• Oscil. damping
• Pressure
Factory setting
Mass flow
Assign diag. beh
Navigation
Expert → Output → Relay output 1 to n → Assign diag. beh (0806–1 to n)
Prerequisite
In the
Relay outp.func.
parameter (→ 149), the
Diag. behavior
option is selected.
Description
Use this function to select the category of the diagnostic events that are displayed for the
relay output.
Selection
• Alarm
• Alarm or warning
• Warning
Factory setting
Alarm
Additional information
Description
If no diagnostic event is pending, the relay output is closed and conductive.
Selection
• Alarm
The relay output signals only diagnostic events in the alarm category.
• Alarm or warning
The relay output signals diagnostic events in the alarm and warning category.
• Warning
The relay output signals only diagnostic events in the warning category.
*
Visibility depends on order options or device settings