Modbus RS485 Register Information
Proline Cubemass 300 Modbus RS485
278
Hauser
Navigation: Expert → Application → Totalizer 1 to n
Parameter
Register
Data type
Access
Selection / User entry / User interface
→
Control Tot. 1 to n (0912–1 to n)
1: 2608
2: 2808
3: 3008
Integer
Read / Write
0 = Totalize
1 = Reset + totalize
2 = hold
3 = Reset + hold
4 = totalize
5 = Hold
Preset value 1 to n (0913–1 to n)
1: 2590 to 2591
2: 2592 to 2593
3: 2594 to 2595
Float
Read / Write
Signed floating-point number
Failure mode (0901–1 to n)
1: 2606
2: 2806
3: 3006
Integer
Read / Write
0 = Stop
1 = Actual value
2 = Last valid value
*
Visibility depends on order options or device settings
6.3.8
"Diagnostics" submenu
Navigation: Expert → Diagnostics
Parameter
Register
Data type
Access
Selection / User entry / User interface
→
Actual diagnos. (0691)
2732
Integer
Read
Symbol for diagnostic behavior,
diagnostic code and short message.
Prev.diagnostics (0690)
2734
Integer
Read
Symbol for diagnostic behavior,
diagnostic code and short message.
Time fr. restart (0653)
2624
Integer
Read
Days (d), hours (h), minutes (m) and
seconds (s)
Operating time (0652)
2631
Integer
Read
Days (d), hours (h), minutes (m) and
seconds (s)
"Diagnostic list" submenu
Navigation: Expert → Diagnostics → Diagnostic list
Parameter
Register
Data type
Access
Selection / User entry / User interface
→
Diagnostics 1 (0692)
2736
Integer
Read
Symbol for diagnostic behavior,
diagnostic code and short message.
Diagnostics 2 (0693)
2738
Integer
Read
Symbol for diagnostic behavior,
diagnostic code and short message.
Diagnostics 3 (0694)
2740
Integer
Read
Symbol for diagnostic behavior,
diagnostic code and short message.
Diagnostics 4 (0695)
2742
Integer
Read
Symbol for diagnostic behavior,
diagnostic code and short message.
Diagnostics 5 (0696)
2744
Integer
Read
Symbol for diagnostic behavior,
diagnostic code and short message.
Содержание Proline Cubemass 300
Страница 2: ......
Страница 294: ...www addresses endress com...