Modbus RS485 Register Information
Proline Cubemass 300 Modbus RS485
250
Hauser
Navigation: Expert → System → Display
Parameter
Register
Data type
Access
Selection / User entry / User interface
→
Value 1 display (0107)
3963
Integer
Read / Write
1 = Mass flow
2 = Volume flow
3 = Correct.vol.flow
4 = Density
5 = Ref.density
7 = Temperature
8 = Osc. ampl. 0
9 = Signal asymmetry
16 = Totalizer 1
17 = Totalizer 2
18 = Totalizer 3
23 = Carr. pipe temp.
32 = Exc. current 0
39 = Electronic temp.
48 = Osc. freq. 0
63 = Osc. damping 0
66 = Pressure
67 = Osc.damp.fluct 0
68 = Freq. fluct. 0
73 = Concentration
74 = Target mass flow
75 = Carrier mass fl.
78 = Carrier vol. fl.
79 = Carr.corr.vol.fl
83 = Oil density
84 = Water density
85 = Density average
86 = GSV flow
87 = GSVa
88 = Oil mass flow
89 = Water mass flow
90 = NSV flow
91 = NSVa
92 = S&W volume flow
93 = Ref.dens.altern.
94 = Oil corr.vol.fl.
95 = Water corr.v.fl.
96 = Targ.corr.vol.fl
97 = Target vol. flow
98 = Temp. average
99 = Oil volume flow
101 = Water cut
102 = Water vol. flow
121 = Curr.output 1
122 = Curr.output 2
0% bargraph 1 (0123)
4136 to 4137
Float
Read / Write
Signed floating-point number
100% bargraph 1 (0125)
4142 to 4143
Float
Read / Write
Signed floating-point number
Decimal places 1 (0095)
3365
Integer
Read / Write
0 = x
1 = x.x
2 = x.xx
3 = x.xxx
4 = x.xxxx