Description of device parameters
Proline Cubemass 300 Modbus RS485
82
Hauser
t
Q
2
3
4
1
A
B
C
D
D
5
6
A0012888
Q
Flow
t
Time
A
Drip
B
Pressure shock
C
Pressure shock suppression active as specified by the time entered
D
Pressure shock suppression inactive
1
Valve closes
2
Flow falls below the on-value of the low flow cut off: pressure shock suppression is activated
3
The time entered has elapsed: pressure shock suppression is deactivated
4
The actual flow value is now displayed and output
5
On value for low flow cut off
6
Off value for low flow cut off
"Partial pipe det" submenu
Navigation
Expert → Sensor → Process param. → Partial pipe det
‣
Partial pipe det
Assign variable (1860)
Low value (1861)
High value (1858)
Response time (1859)
Max. damping (6040)
Содержание Proline Cubemass 300