Proline Cubemass 300 Modbus RS485
Description of device parameters
Hauser
185
Additional information
Display
Via the local display: the time stamp and corrective measures referring to the cause of
the diagnostic message can be accessed via the key.
Example
For the display format:
F271 Main electronics
Timestamp
Navigation
Expert → Diagnostics → Timestamp
Description
Displays the operating time when the last diagnostic message before the current message
occurred.
User interface
Days (d), hours (h), minutes (m) and seconds (s)
Additional information
Display
The diagnostic message can be viewed via the
Prev.diagnostics
parameter
Example
For the display format:
24d12h13m00s
Time fr. restart
Navigation
Expert → Diagnostics → Time fr. restart (0653)
Description
Use this function to display the time the device has been in operation since the last device
restart.
User interface
Days (d), hours (h), minutes (m) and seconds (s)
Operating time
Navigation
Expert → Diagnostics → Operating time (0652)
Description
Use this function to display the length of time the device has been in operation.
User interface
Days (d), hours (h), minutes (m) and seconds (s)