Description of device parameters
Proline Cubemass 300 Modbus RS485
104
Hauser
Failure mode (1601–1 to n)
Failure value (1602–1 to n)
Terminal no.
Navigation
Expert → Input → Current input 1 to n → Terminal no. (1611–1 to n)
Description
Displays the terminal numbers used by the current input module.
User interface
• Not used
• 24-25 (I/O 2)
• 22-23 (I/O 3)
Additional information
"Not used" option
The current input module does not use any terminal numbers.
Signal mode
Navigation
Expert → Input → Current input 1 to n → Signal mode (1610–1 to n)
Prerequisite
The measuring device is
not
approved for use in the hazardous area with type of
protection Ex-i.
Description
Use this function to select the signal mode for the current input.
Selection
• Passive
• Active
Factory setting
Active
Current span
Navigation
Expert → Input → Current input 1 to n → Current span (1605–1 to n)
Description
Use this function to select the current range for the process value output and the upper
and lower level for signal on alarm.
Selection
• 4...20 mA
• 4...20 mA NAMUR
• 4...20 mA US
• 0...20 mA