Description of device parameters
Proline Cubemass 300 Modbus RS485
86
Hauser
Selection
• Liquid
• Gas
Factory setting
Liquid
Select gas type
Navigation
Expert → Sensor → Measurement mode → Select gas type (6074)
Prerequisite
The
Gas
option is selected in the
Select medium
parameter (→ 85).
Description
Use this function to select the type of gas for the measuring application.
Selection
• Air
• Ammonia NH3
• Argon Ar
• Sulf. hex.fl.SF6
• Oxygen O2
• Ozone O3
• Nitrog. ox. NOx
• Nitrogen N2
• Nitrous ox. N2O
• Methane CH4
• Hydrogen H2
• Helium He
• Hydrog.chlor.HCl
• Hydrog.sulf. H2S
• Ethylene C2H4
• Carbon diox. CO2
• Carbon monox. CO
• Chlorine Cl2
• Butane C4H10
• Propane C3H8
• Propylene C3H6
• Ethane C2H6
• Others
Factory setting
Methane CH4
Additional information
Description
The gas type needs to be selected so that it is possible to comply with accuracy
specifications in gas applications.
Sound velocity
Navigation
Expert → Sensor → Measurement mode → Sound velocity (6147)
Prerequisite
In the
Select gas type
Others
option is selected.
Description
Use this function to enter the sound velocity of the gas at 0 °C (+32 °F).