
18 WGFMU
18.7
WGFMU Pattern Editor (A.03.20)
[Supported Analyzer]
B1500A
[Description]
Performs the WGFMU channel output setup and the measurement setup.
Set the ExecutionMode to Run Vector and click the Single button to perform the WGFMU channel output and
measurement.
[Test Parameters]
ExecutionMode: Execution mode
Run Vector: Performs the waveform output and measurement
Pattern Validation: Displays the waveform and measurement setup, for debugging
[WGFMU1 and WGFMU2]
Enable: Enable (uses the channel) or Disable (does not use)
Channel: WGFMU channel to use
OperationMode: Operation mode
PG Vmeas (PG mode, voltage measurement)
Fast IV Imeas (Fast IV mode, current measurement)
Fast IV Vmeas (Fast IV mode, voltage measurement)
VForceRange: Voltage output range
Auto, 3V, 5V, -10V to 0V, or 0V to 10V
InitMeasRange: Measurement range
For current measurement: 1uA, 10uA, 100uA, 1mA, or 10mA
For voltage measurement: 5V or 10V
[Pattern]
RepeatCount: Repeat count of waveform data and measurement event data
WaveformCh1: Channel1 waveform data
WaveformCh2: Channel2 waveform data
MeasurementEvent: Measurement event data
Waveform data and measurement event data can be defined by using the Define vector data dialog box. To open the
Define vector data dialog box, click the left button of each field. Clicking the * button on the dialog box increases the
row.
Waveform data:
Column 1: Time (absolute value)
Column 2: Voltage output value
Measurement event data:
Column 1: Sampling measurement start time (absolute value)
Column 2: Number of measurement points per one sampling
Column 3: Sampling interval
Column 4: Averaging time per one point measurement
Column 5: Ch1 range event
Column 6: Ch2 range event
For the column 5 and 6, 0 is set normally. For the current measurement, setting a number from 1 to 5 enables the
range event. This changes the current measurement range to the specified range when the sampling measurement is
started.
1: 1uA, 2: 10uA, 3: 100uA, 4: 1mA, 5: 10mA
To set the range event only without the measurement event, enter 0 to the column 2 and set the column 5 and 6.
Agilent EasyEXPERT Application Library Reference, Edition 8
18-13
Содержание EasyEXPERT
Страница 1: ...Agilent Technologies Agilent EasyEXPERT Application Library Reference ...
Страница 17: ...1 BJT ...
Страница 63: ...2 CMOS ...
Страница 119: ...3 Discrete ...
Страница 127: ...4 GenericTest ...
Страница 133: ...5 MCSMU_IV ...
Страница 141: ...6 Memory ...
Страница 211: ...7 MixedSignal ...
Страница 249: ...8 NanoTech ...
Страница 265: ...9 Organic ...
Страница 269: ...10 PwrDevice ...
Страница 285: ...11 Reliability ...
Страница 377: ...12 Sample ...
Страница 381: ...13 Solar Cell ...
Страница 401: ...14 SPGU_PLSDIV ...
Страница 409: ...15 Structure ...
Страница 459: ...16 TFT ...
Страница 463: ...17 Utility ...
Страница 478: ...17 Utility Agilent EasyEXPERT Application Library Reference Edition 8 17 16 ...
Страница 479: ...18 WGFMU ...
Страница 493: ...19 WGFMU_IV ...
Страница 497: ...19 WGFMU_IV Id Drain current Vg gate voltage Agilent EasyEXPERT Application Library Reference Edition 8 19 5 ...
Страница 501: ...19 WGFMU_IV Id Drain current Vd Drain voltage Agilent EasyEXPERT Application Library Reference Edition 8 19 9 ...
Страница 505: ...20 GaN Diode ...
Страница 511: ...21 GaN FET ...
Страница 523: ...22 IGBT ...
Страница 549: ...23 Interconnection ...
Страница 553: ...24 MISCAP ...
Страница 561: ...25 PowerBJT ...
Страница 576: ...25 PowerBJT Agilent EasyEXPERT Application Library Reference Edition 8 25 16 ...
Страница 577: ...26 PowerDiode ...
Страница 585: ...27 PowerMOSFET PMIC SiC ...