9-20
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Service and Maintenance
Weekly Maintenance Procedures
Section 9
Figure 9.6 Sample Transfer Peristaltic Pump
To check the peristaltic pump tubing, follow the procedure below:
1. In the
MAIN MENU
, press
[SPECIAL PROTOCOLS]
followed by
>
DIS/ENAB
ANALYZER]
and
[DISABLE ANALYZER]
to disable the Analyzer.
2. Gently pull on the hand hold of the Left Front Cover to swing the cover open.
3. Locate the Sample Transfer Pump on the left side of the Flow Panel. (Refer
to Figure 9.6.)
4. Push the Pump Shoe away from the pump wheel and slip the tubing out.
(Refer to Figure 9.6.)
5. Visually inspect the tubing for damage. If the tubing is damaged or flattened,
change the tubing. (Refer to
Peristaltic Pump Tubing Replacement
later in
this section.)
6. Using your fingers, massage the tubing to test its strength and resiliency and
to remove any indentation caused by the wheel.
7. Push the Pump Shoe away from the pump wheel and slip the tubing back
under the wheel. Make sure the tubing is securely positioned under the wheel.
8. Close the Left Front Cover and press into place.
9. Press
[ENABLE ANALYZER]
to return the instrument to the READY state.
10. Press
[MAIN]
to return to the
MAIN MENU
.
Extended Auto-Clean
The Extended Auto-Clean cycle (refer to the Monthly Procedures later in this
section) should be run weekly when routinely running reticulocytes.
1 Tubing
2 Pump Shoe
3 Pump Wheel
1
2
3
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...