CELL-DYN
®
3200 System Operator’s Manual
10-87
9140181H—October 2001
Section 10
Troubleshooting and Diagnostics
RBC Flow Error
/BL
RBC FLOW ERROR
is displayed on the RUN screen to the right of the RDW result.
“
RBC FLOW ER
” is printed on the graphics report to the right of the RDW result.
Results for RBC, PLT, and related parameters are suppressed.
– – – - Probable Cause – – – –
– – – - Corrective Action – – –
1. An increasing RBC count rate was
detected in the optical flow cell
during the RBC measurement.
1. Repeat the sample.
2. Optical flow cell is dirty or contains
bubbles.
2. Empty/refill flow cell in Special
Protocols
Run AutoClean and/or Extended
AutoClean
3. Sample Dilution Delivery Problem.
3. Check/replace tubing in Sample
Transfer Pump.
Clean Sample Injection Syringe,
check for leakage, and replace as
necessary.
4. Diluent/Sheath flow (partially)
obstructed.
4. Replace Diluent/Sheath Filter.
3 consecutive WOC flow errors
/BL
3 consecutive NOC flow errors
/BL
3 consecutive RBC flow errors
/BL
– – – - Probable Cause – – – –
– – – - Corrective Action – – –
1. Three consecutive flow error
messages occurred during Sample
Loader operation.
1. Three consecutive flow errors of the
same type will cause the Sample
Loader to halt. Refer to the three
previous messages and
Troubleshooting
Tips and
Techniques
earlier in this section.
2. If unable to resolve this problem,
contact your Abbott Hematology
Customer Support Center.
Table 10.3 Sample-Related Fault Conditions
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...