Performance Specifications
Background Counts
Background counts for the following parameters are acceptable up to the limits
listed:
WOC <0.10 K/µL
NOC
<0.10 K/µL
RBC
<0.02 M/µL
HGB
<0.10 g/dL
PLT
<5.0 K/µL
RETIC <100 counts/count cycle
Carryover
Carryover is the degree of influence the previous sample has on the next sample
and is expressed as a percent. Fresh whole blood samples that are used to verify
carryover specifications should have results that fall within each laboratory’s
reference interval (normal range). Each sample is run in triplicate followed by three
background cycles.
The percent carryover for CBC parameters, as shown in Table 4.6, is calculated
using the following formula:
Absolute Carryover = Background
1
- Background
3
Table 4.6
Carryover for WOC, NOC, RBC, HGB, and PLT
Parameter
Specimen Range
Carryover
WOC
4.1–11.0 x K/
µL
< 1% or 0.1 K/
µL
NOC
4.1–11.0 x K/
µL
< 1% or 0.1 K/
µL
RBC
4.0–6.0 x M/
µL
< 1% or 0.03 M/
µL
HGB
12.1–17.0 g/
dL
< 1% or 0.1 g/
dL
PLT
150.0–400.0 x K/
µL
< 1% or 6 K/
µL
% Carryover =
Background
1
-
Background
3
Background
3
Sample
3
-
X 100
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...