CELL-DYN
®
3200 System Operator’s Manual
6-45
9140181H—October 2001
Section 6
Calibration Procedures
Determining Which Parameters Need Calibration
To determine which parameters require calibration in the Closed Mode, follow the
procedure below:
1. Obtain the values derived in column 4 of Worksheet 4 and enter these values
in column 1 of Worksheet 5, Mode to Mode Calibration Criteria, provided in
subsection
Manual Calibration Worksheets
at the end of this section.
NOTE:
Make copies of this worksheet as needed.
2. Follow the instructions on Worksheet 5 to determine which parameters
require calibration.
NOTE:
Make copies of this worksheet as needed.
3. For each parameter, if the % Diff is
equal to or less than
the value in the
Lower Limit column, then CALIBRATION IS NOT REQUIRED for that
parameter because the value is within range.
a. If none of the parameters requires Closed Mode calibration and Open
Mode has not been calibrated, go to
Calibration Procedure — Open
Mode
to determine if Open Mode needs calibrating.
NOTE:
It is recommended that the primary mode (Open Mode) be
calibrated first.
b. If none of the parameters requires Closed Mode calibration and Open
Mode calibration has already been completed, press
[MAIN]
to return to
the
MAIN MENU
.
4. For each parameter, if the % Diff falls
between
the upper and lower
calibration range, shown in the Calibration Range column, then
CALIBRATION IS REQUIRED.
5. For each parameter, if the % Diff is
greater than
the value in the Upper Limit
column, there may be a computation error or an instrument problem. In this
case, do the following:
a. Recheck all numbers and calculations on Worksheets 4 and 5.
b. Determine if any component on the instrument was changed. This could
affect calibration. Such components include the Shear Valve, Optical
Flow Cell, Hemoglobin Flow Cell, or one of the syringes.
c. If a component was changed, then treat the result as if it fell within the
“calibration range” (even though it is greater than the upper limit).
CALIBRATION IS REQUIRED for that parameter.
d. If no component was changed and your calculations are correct, DO
NOT CALIBRATE. Confirm that all Pre-Calibration procedures were
completed and call the Abbott Hematology Customer Support Center for
assistance (at 1-800-CELL-DYN in the U.S.).
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...