Troubleshooting and Diagnostics
Tables for Instrument Conditions and Messages
Section 10
Tower Fault 141: invalid tube height
/BL (Continued)
b. A mechanical problem is
preventing the tube spinning
assembly with tube height sensor
flag from moving up and down
properly.
b. Remove the Loader Cover.
Check for an obstruction that is
preventing the tube spinning
assembly with tube height sensor
flag from moving up and down
properly and, if one is found,
remove it. Then, reinstall the
Loader Cover, reset the racks,
and reset the Sample Loader.
c. The tube height sensor flag is
bent or the tower sensors, motor,
or related electronics are
defective.
c. Contact your Abbott
Hematology Customer Support
Center.
Tower Fault 142: no tube present
/BL
The tube height sensor did not sense a tube when one was expected.
– – – - Probable Cause – – –
– – – - Corrective Action –
1. The tube being processed is too
short.
1. To run the sample in the Sample
Loader, pour the sample into a tube
without anti-coagulant that meets
height specifications. Then, reset
the racks and rest the Sample
Loader by pressing the following
keys in order:
[CLEAR FAULT]
,
[START LOADER]
,
[RESET LOADER]
.
NOTE:
For tube height
specifications, refer to
Physical and
Performance
Specifications
2. A failure of the sensor or related
electronics has occurred.
2. Reset the racks and reset the
Sample Loader.
Table 10.2 General Fault Conditions (Continued)
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...