CELL-DYN
®
3200 System Operator’s Manual
3-33
9140181K—July 2002
Section 3
Principles of Operation
Patient, QC, Resistant RBC Mode Flagging (when WOC and NOC are equivalent)
Descriptor/Flag
Cause
Suggested Action
BAND
1) The BAND flag is triggered if any of
the following conditions are met:
2) The CV% of the neutrophil cluster on
the 0
°
axis exceeds expected criteria.
3) The instrument detects > 12.5% of the
total WBC count as BAND cells
The ratio of suspected bands to mature
neutrophils is >50%
Review a stained smear for the presence of
bands and follow your laboratory’s review
criteria.
NOTE: When bands are present, they are
included in the total neutrophil count.
IG
The IG flag is triggered if the following
condition is met:
The instrument detects > or = 3% of the
total WBC count as IG cells
Review a stained smear for the presence of
immature granulocytes and follow your
laboratory’s review criteria.
NOTE: When IGs are present, they are
included in the total neutrophil count.
BLAST
The BLAST flag is triggered if the
following condition is met:
The instrument detects > 1% of the total
WBC count as Blast cells
Review a stained smear for the presence of
blasts and follow your laboratory’s review
criteria.
NOTE: When blasts are present, they are
included in the monocyte count.
VAR LYM
When the FWBC flag is triggered, the
VAR LYM flag is always set.
In addition, the flag is set if any of the
following attributes apply:
1) Position of the lymphocyte cluster on
the scatter plot is abnormal.
2) The ratio between lymphocytes and
other WBC subpopulations is increased.
3) Lymphocyte count (absolute or %) is
elevated
Review a stained smear for the presence of
variant lymphocytes and follow your
laboratory’s review criteria.
NOTE: When variant lymphocytes are
present, they are included in the
lymphocyte count.
NOTE: This flag may be displayed singly or in
combination with the blast flag. If the
flag is displayed with the blast flag, it
is displayed as VLYM/BLAST.
RBC MORPH
One or more of the following parameters
exceeds expected limits:
MCV < 80fL or >100fL
MCH < 25pg or >34pg
MCHC < 29g/dL or >37g/dL
RDW >18.5%
Review a stained smear for abnormal RBC or
PLT morphology and follow your laboratory's
review criteria.
Table 3.4
Specimen Modes (Continued)
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...