Section 10
Troubleshooting and Diagnostics
Tower Fault 141: invalid tube height
/BL
The tube height sensor system indicates that the tube at the Aspiration Station
exceeds height specification.
– – – - Probable Cause – – –
– – – - Corrective Action –
1. The instrument cannot
accommodate the height of the tube.
1. To run the sample in the Sample
Loader, pour the sample into a tube
without anti-coagulant that meets
height specification. Then reset the
racks and reset the Sample Loader
by pressing the following keys in
order:
[CLEAR FAULT]
,
[START LOADER]
,
[RESET LOADER]
.
NOTE:
For tube height
specifications, refer to
Physical and
Performance
Specifications
.
2. Failure of the sensor or related
electronics has occurred.
2. Reset the racks and reset the
Sample Loader.
3. The tube height sensor or sensor flag
on the Tower is not moving into
position properly.
a. The Guide Shafts are dirty.
3. The tube height sensor or sensor
flag on the Tower is not moving
into position properly.
a. Remove the Loader Cover.
Using a DYN-A-WIPE pad or
lint-free wipe moistened with
isopropyl alcohol, clean the
Guide Shafts. Then, reinstall the
Loader cover, reset the racks,
and reset the Sample Loader.
NOTE:
The Guide Shafts are the
three vertical bars on the
tower.
Table 10.2 General Fault Conditions (Continued)
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...