CELL-DYN
®
3200 System Operator’s Manual
10-61
9140181H—October 2001
Section 10
Troubleshooting and Diagnostics
Tower/Loader Fault 19: invalid process command
/BL
The Analyzer command given to the module that controls the tower and loader
is invalid.
– – – - Probable Cause – – –
– – – - Corrective Action –
1. A software error has occurred.
1. Reboot the System.
NOTE:
If you wish to use the
Sample Loader, reset the
racks before rebooting the
System.
2. If the message appears repeatedly,
contact your Abbott Hematology
Customer Support Center.
Loader Fault 40: mix head stuck in angle position
/BL
The Mix Head failed to return to its vertical position.
– – – - Probable Cause – – –
– – – - Corrective Action –
1. A mechanical problem is preventing
the Mix Head from rotating
downward.
1. Remove the Loader Cover. Check
for an obstruction that is preventing
the Mix Head from rotating
downward. If an obstruction is
found, remove it and rotate the Mix
Head to its vertical position.
Reinstall the Loader Cover, reset
the racks, and reset the Sample
Loader by pressing the following
keys in order:
[CLEAR FAULT]
,
[START LOADER]
,
[RESET LOADER]
.
2. A failure of the magnetic reed sensor
or related electronics has occurred.
2. If the message appears repeatedly,
contact your Abbott Hematology
Customer Support Center.
Table 10.2 General Fault Conditions (Continued)
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...