CELL-DYN
®
3200 System Operator’s Manual
6-21
9140181K—July 2002
Section 6
Calibration Procedures
Auto-Cal Menu
When the
[AUTO-CALIBRATE]
key is pressed in the main
CALIBRATION
menu, the
AUTO-CALIBRATION
menu is displayed (refer to Figure 6.6). Reference values are
entered on this screen.
The Spec ID field in the upper left corner of the screen defaults to “CAL#01” to
identify the sample as a “Calibration sample.” This field is user-definable and
accepts up to 6 alphanumeric characters. As each calibration sample is run, the
calibration program adds a “-” and the run number from “01” to “10.”
Figure 6.6 Auto Calibration Screen
Start Auto-Cal
Figure 6.6 shows the
AUTO CALIBRATION
screen before any reference values are
entered. When the first reference value has been entered and the Enter key pressed,
the
[START AUTO CAL]
and
[QUIT]
keys are displayed (refer to Figure 6.7).
After the
[START AUTO CAL]
key is pressed, the operator will not be able to toggle
between Calibrator and Whole Blood nor make changes to the reference values, the
Spec ID, or the Operator ID. The instrument is ready to begin processing samples.
NOTE:
Pressing the
[WHOLE BLOOD]
or
[CALIBRATOR]
key does not clear or
change the reference values.
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...