CELL-DYN
®
3200 System Operator’s Manual
1-11
9140181H—October 2001
Section 1
Use or Function
Syringe Assembly
There are two syringe driver assemblies, each containing two syringes. Each
syringe is operated by its own stepper motor. The function of each syringe is
described below:
•
Diluent/Sheath Syringe
— (1) delivers a specific volume of diluent to
transport the RBC segment from the Shear Valve to the RBC/PLT Mixing
Chamber and to dilute the segment prior to measurement, and (2) delivers a
specific volume of diluent to transport the HGB segment from the Shear
Valve to the HGB Mixing Chamber and to dilute the segment prior to
measurement.
•
HGB Lyse Syringe
— delivers a specific volume of HGB Lyse to the HGB
Mixing Chamber/Flow Cell to further dilute the HGB segment prior to
measurement.
•
WBC Lyse Syringe
— delivers a specific volume of WBC Lyse to transport
the WBC segment from the Shear Valve to the WBC Mixing Chamber and to
dilute the segment prior to measurement.
•
Sample Injection Syringe
— injects a specific volume of the diluted sample
into the Optical Flow Cell for RBC/PLT, WBC (WOC), and WBC (NOC)
measurements.
Sample Transfer Peristaltic Pump
The
Sample Transfer Peristaltic Pump
is composed of a rotor and a pump tube
holder. It is used to transfer the WBC dilution, RBC/PLT dilution, and HGB/NOC
dilution to the Optical Flow Cell from their respective mixing chambers.
HGB Flow Cell and Mixing Chamber
The
HGB Flow Cell Assembly
is integrated with a mixing chamber and contains
the following components:
• A fully enclosed (light-tight) mixing chamber with optical windows and
electronics
• An
LED Light Source
• A
Photodetector
for measuring the light transmitted.
WBC Mixing Chamber
The
WBC Mixing Chamber
is used to swirl mix the WBC dilution.
RBC/PLT Mixing Chamber
The
RBC/PLT Mixing Chamber
is used to swirl mix the RBC/PLT dilution.
Wash Block
There are two
Wash Blocks
on the instrument, one for the Open Sample Aspiration
Probe and one for the Closed Sample Needle. Both Wash Blocks work in a similar
fashion: After aspiration, they move down the outside of the probe or needle,
rinsing it with diluent. The diluent rinse is then transferred to the waste chamber.
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...