CELL-DYN
®
3200 System Operator’s Manual
3-31
9140181K—July 2002
Section 3
Principles of Operation
Data Flagging
Table 3.4
Specimen Modes
Patient, QC, Resistant RBC Mode Flagging (when WOC and NOC are equivalent)
Descriptor/Flag
Cause
Suggested Action
NWBC
1) When stroma interference is >1.5% of
analyzed events and there is no
declining WOC kinetic rate.
2) In patients with leukopenia the NWBC
is set when the stroma interference is
>2.5.%
Review smear for platelet clumps, giant
platelets or low levels of NRBCs and follow
your laboratory’s review criteria.
If no other suspect parameter flags are
present, the WBC and differential may be
reported.
NRBC
When stroma interference is >3.0% of
analyzed events and there is no
declining WOC kinetic rate,
or
When the patient is anemic and the
stroma interference is >0.9% of
analyzed events and there is no
declining WOC kinetic rate.
Review smear for the presence of NRBCs and
follow your laboratory’s review criteria.
NRBC
WBC
DFLT (NLMEB)
In addition to the criteria above, if the
number of analyzed events is high or
indicates a likely interference in the WBC
count, then the WBC alert and DFLT
NLMEB flags are also given.
Review smear for the presence of NRBCs and
follow your laboratory’s review criteria.
Confirm WBC count by alternate method.
RRBC
WBC
DFLT (NLMEB)
Stroma interference is high and a
declining WOC kinetic rate is detected.
Repeat in Resistant RBC Mode.
If flag persists, review for presence of NRBCs
and verify Lymph valve. Confirm WBC count
by alternate method.
FWBC
VAR LYMPH
WBC
DFLT (NLMBEB)
Stroma interference is low but a
declining WOC kinetic rate is detected,
or
Stroma interference is low and there is
no declining WOC kinetic rate, but
WOC>8.0 K/uL and LYM%>80%.
Repeat in FWBC mode. NOC result will be
reported as WBC result.
Review smear to confirm lymph count and
presence of fragile WBCs.
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...