![Abbott CELL-DYN 3200 Скачать руководство пользователя страница 403](http://html1.mh-extra.com/html/abbott/cell-dyn-3200/cell-dyn-3200_system-operators-manual_450985403.webp)
CELL-DYN
®
3200 System Operator’s Manual
9-33
9140181H—October 2001
Section 9
Service and Maintenance
14. Align the lock notch of the center section with the mounting guide. (Lock
notch faces the mounting guide, smaller rim notch faces down.) Carefully
slide this section back onto the mounting arm until it is flush with the rear
section. (Refer to Figure 9.10.)
CAUTION:
Results may be affected if the center section is installed back-
wards.
Be certain
the rim notch faces down.
15. Wipe the inner surface of the front section with a DYN-A-WIPE
(or other
lint-free pad) dampened with warm water. Use care to avoid scratching the
inner surface. Do not dry.
NOTE:
Hold the section by the edges to avoid getting fingerprints on the
inner surfaces.
16. Align the lock notch of the front section with the mounting guide. A pin on
the inner rim should align with the groove on the mounting arm. Carefully
slide this section back until it touches the center section.
17. Firmly hold the three valve sections together and replace the Shear Valve
Retaining Screw. Turn the screw clockwise until it stops.
18. Remove any cleaning materials (such as gauze pads) that may have been left
on the instrument.
19. Press
[RESTORE SHEAR VAL]
WRUHWXUQWKH6KHDU9DOYHWRLWVRSHUDWLQJ
SRVLWLRQDQGWRUHWXUQWKHLQVWUXPHQW
to the READY state.
20. Replace the Tower Cover (Tower Door on the CS model must be open).
21. Press
[RETURN]
followed by
[MAIN]
DQG
[RUN]
to display the
RUN
screen.
22. Run at least five background counts to rinse the system. Check that the
background counts are acceptable before running controls or patient samples.
If the counts are unacceptable, troubleshoot accordingly (refer to
Troubleshooting and Diagnostics
23. Record this maintenance in your maintenance log.
Open Sample Probe Interior Cleaning
The Open Sample Aspiration Probe is thoroughly cleaned whenever the Auto-
Clean cycle is performed. However, additional cleaning may be required from time
to time. The probe may be cleaned manually, as described below.
Materials Required
1. Gloves, lab coat, and safety glasses
2. Syringe (10 cc or larger) with at least 3" of 1/32" silicone tubing attached to
the tip
3. Small needle-nose pliers or similar tool
4. Small beaker or container
5. Deionized water
6. Cleaning solution (0.5% sodium hypochlorite solution). See the formula for
mixing this solution under
Decontamination Procedures
earlier in this
section.
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...