9-48
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Service and Maintenance
Nonscheduled Maintenance Procedures
Section 9
Reagent Line Cleaning
The reagent lines should be cleaned if contamination is suspected. To clean the
lines, refer to steps 1–4 in
Preparation for Shipping or Extended Period of
Non-Use
later in this section.
Open Sample Probe Replacement
Materials Required
1. Gloves, lab coat, and safety glasses
2. Replacement probe
3. 7/64" Allen wrench
4. Small needle-nose pliers or similar tool
Procedure
1. In the
MAIN MENU
, press
[SPECIAL PROTOCOLS]
IROORZHGE\
[DIS/ENAB
ANALYZER]
and
[DISABLE ANALYZER]
.
2. Gently pull on the hand hold of the Right Front Cover and swing the cover
open to access the right side of the Flow Panel. Remove Tower Cover.
3. Locate the tubing attached to the top of the Open Sample Probe. (Refer to
Figure 9.19.) Hold the probe firmly with one hand and with the other hand
use the pliers to carefully work the tubing up and off the top of the probe.
4. Using the Allen wrench, remove the two hex nuts holding the Probe Bracket
Arm to the Bracket Support Arm on the Probe Assembly Frame.
5. Pull the probe up and out of the Wash Block.
6. Insert the new probe into the Wash Block.
7. Place the Probe Bracket Arm on top of the Bracket Support Arm. Align the
holes in the bracket arm to the holes in the Probe Assembly Frame. Insert and
tighten the two hex nuts using the Allen wrench.
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...