The Sample Loader beeps but does not begin processing
– – – - Probable Cause – – – –
– – – - Corrective Action – – –
1. The internal cable that connects the
Sample Loader to the Analyzer is
loose or disconnected.
1. Contact your Abbott Hematology
Customer Support Center.
2. Circuitry malfunction.
2. Press
[FAULT REPORT]
in the main
DIAGNOSTICS MENU
and record any
fault listed.
3. Turn the instrument OFF and ON
again and prime the instrument.
4. If the problem persists, contact your
Abbott Hematology Customer
Support Center.
The Display Monitor membrane touch keys are not operational
– – – - Probable Cause – – – –
– – – - Corrective Action – – –
1. The computer is performing a
function that inhibits the keys.
1. No action required. Wait for the
computer to resume the operation.
Refer to the screen for the current
Status Box message.
2. There is an incomplete operator
entry.
2. Complete the operator entry or press
the ESC key on PC keyboard.
3. A data transmission to the printer or
laboratory computer is in process.
3. No action required. Wait until the
transmission is completed.
4. Membrane keyboard or circuitry
malfunction.
4. Power OFF the Analyzer. Then
reseat the Membrane Keypad cable
on the rear panel of the Data Module
and power Analyzer ON.
NOTE:
For the location of the
Membrane Keypad
Interface connector, refer to
.
Table 10.4 Non-Functional Fault Conditions (Continued)
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...