![Abbott CELL-DYN 3200 Скачать руководство пользователя страница 364](http://html1.mh-extra.com/html/abbott/cell-dyn-3200/cell-dyn-3200_system-operators-manual_450985364.webp)
8-4
CELL-DYN
®
3200 System Operator’s Manual
9140181K—July 2002
Hazards
Hazard Information and Precautions
Section 8
Chemical Hazards
Prevent exposure to chemicals used in the operation and maintenance of the
CELL-DYN 3200 System (including reagents) by using appropriate personal
protective equipment, work procedures, and information on Material Safety Data
Sheets (MSDS). Refer to
Installation Procedures and Special
for an installation procedure for chemical containers.
Electrical Hazards
Basic electrical hazard awareness is essential to the safe operation of any
hematology analyzer. To ensure safe operation of the CELL-DYN 3200 System:
• Periodically inspect electrical cabling into and on the instrument for signs of
wear or damage.
• When moving equipment, lift all power cables clear of all system components.
CAUTION: Electrical Hazard.
Do not disconnect any electrical connec-
tion while the power is on. Follow instructions for correctly powering down
the instrument and all connected equipment before performing maintenance
on parts which require protective covers to be removed for access. Use only
approved power cords and electrical accessories, supplied with the instru-
ment, or provided by Abbott, to protect against electrical shock.
CAUTION: Electrical Hazard.
Turn off the power to the instrument and
disconnect the power cord before removing any instrument panel that is
securely fastened in place by screws or prior to replacing fuses. Replace
only the externally accessible fuse located immediately above the power
cord connector on the rear panel of the instrument. Use replacement fuses
only of the specified type and electrical rating.
• Keep liquids away from all electrical connectors (such as electrical outlets)
or communication connectors (such as the LIS connector).
• Keep the floor dry.
• The electrical circuit spacing of the CELL-DYN 3200 System is based on
pollution degree (1) and altitude [up to 2000 M (6500 ft)] as per IEC
61010-1. Pollution degree 1 is defined as an environment where there is no
pollution or only dry, non-conductive pollution.
CAUTION:
If the instrument is used or modified in a manner not specified
by the manufacturer, the protection provided by the instrument may be
impaired.
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...