CELL-DYN
®
3200 System Operator’s Manual
5-119
9140181H—October 2001
Section 5
Operating Instructions
4. Place the well-mixed sample under the Open Sample Aspiration Probe, make
sure the end of the probe is deeply immersed in the sample, and press the
Touch Plate to start the cycle.
NOTE:
Additional entries may be made to the Work List while processing
takes place. To input additional entries to the Work List, press
[WORK LIST]
to display the
WORK LIST
screen. Type in the new
entries.
5. When all samples have been processed, check to see if all entries have been
deleted from the Work List. Entries are automatically deleted from the Work
List when a match occurs. Entries remaining in the Work List indicate a
successful match did not occur. The operator should follow laboratory
procedures to determine the cause of this discrepancy.
Running STAT Samples with Bar Codes
STAT samples with bar codes are processed in the same manner as described in the
sections
Using the Work List with Bar Codes
and
Running Samples
immediately
above.
Using the Work List with Specimen IDs
If Work List records with Specimen IDs (non bar code) were downloaded from a
host, the operator can use the Work List by entering the Specimen ID on the sample
tube into the <Next ID> field on the
RUN
screen. The instrument will attempt to
match the number in the <Next ID> field on the
RUN
screen with the same number
in the <Specimen ID> field in the Work List.
If a match occurs and the sample is processed, the results along with the
demographic data from the Work List will be displayed on the
RUN
screen and sent
to the Data Log. That entry will be deleted from the Work List.
If a match does not occur and the sample is processed, only the sample results and
Specimen ID entered on the
RUN
screen will be displayed on the
RUN
screen and
stored in the Data Log. No entry will be deleted from the Work List. The operator
must go to the Data Log to enter the remaining demographic information.
Running Samples
1. In the
RUN
screen, press
[SPECIMEN TYPE]
followed by the appropriate
specimen type (Patient, Fragile WBC, or Resistant RBC).
2. If necessary, press
[CHANGE SAMPLER]
to select Open Mode.
3. Enter the Specimen ID into the <Next ID> field on the
RUN
screen.
4. Place the well-mixed sample under the Open Sample Aspiration Probe, make
sure the end of the probe is deeply immersed in the sample, and press the
Touch Plate to start the cycle.
NOTE:
Additional entries may be made to the Work List while processing
takes place. To input additional entries to the Work List, press
[WORK LIST]
to display the
WORK LIST
screen. Type in the new
entries.
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...