2-6
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Installation Procedures and Special Requirements
Tubing Installation
Section 2
Figure 2.1 Reagent Inlet Panel
7. Locate the
Waste-Full Sensor
Plug attached to the cap's electrode wires.
Insert the plug into the
:DVWH6HQVRU
connector located on the Reagent Inlet
Panel. Attach the ground shield on the cable to the connection on the rear
panel.
If the waste tube is placed directly into a drain, insert the Dummy Plug
(provided in the Accessory kit) into the Waste Sensor Connector, as the
EXTERNAL WASTE FULL
alert will activate if no plug is inserted. For
information on ordering a Waste Dummy Plug, refer to
Appendix B
¥
Parts
and Accessories
.
Normally Closed Valves
To gain access to all six Normally Closed Valves on the Flow Panel, open both the
Left and Right Front Covers and remove the Tower Cover. Remove the Front Skirt
(CS model) or move the Sample Loader (SL model) away from the instrument.
(Refer to instructions later in this section for opening and removing these
components.)
Before shipment, the tubing is removed from these valves. To ensure correct
system operation, the tubing must be completely inserted in all the valves before
the instrument power is turned ON. Follow the directions below to open the front
covers and remove the Skirt or Sample Loader to allow access to all the Normally
Closed Valves. Refer to Figure 2.2.
1 Waste Full Sensor
2 Waste Outlet
Connector
3 HGB Lyse Connector
4 Diluent/Sheath
Connector
5 WBC Lyse Connector
1
2
3
4
5
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...