CELL-DYN
®
3200 System Operator’s Manual
6-47
9140181H—October 2001
Section 6
Calibration Procedures
Entering New Calibration Factors — Closed Mode
1. In the
CALIBRATION
menu, press
[ENTER FACTOR]
.
2. Use the arrow keys to select the first factor in the Closed Sampler column to
be changed. Type in the New Calibration Factor calculated in
Calculating
New Closed Mode Calibration Factors
above. Press the Enter key to save the
factor and advance the cursor. If necessary, use the arrow keys to select the
next factor to be changed.
NOTE:
Prior to exiting the screen, the operator may press the
[RESTORE
FACTORS]
soft key to recall factors, stored on the Hard Disk,
corresponding to the current mode — Open or Closed.
[RESET ALL
TO 1.000]
is used to reset all factors displayed on the screen to
1.000.
3. When all new factors have been entered, press
[RETURN]
to display the
CALIBRATION LOG
screen with the new calibration factors.
NOTE:
If no changes are made in the
ENTER CALIBRATION FACTOR
screen,
pressing
[RETURN]
will bypass the
CALIBRATION LOG
screen and
display the main C
ALIBRATION
menu.
4. Complete the Calibration Log (adding comments) as required. If a printout of
the log is desired, press
[PRINT LOG]
.
5. Press
[RETURN]
twice to return to the main
CALIBRATION
menu.
6. Press
[PRINT]
to obtain a copy of the new Closed Sampler Calibration Factors.
7. Press
[MAIN]
to return to the
MAIN MENU
.
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...