6-8
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Calibration Procedures
Pre-Calibration Procedures
Section 6
10. Confirm that the waste container is no more than half full. If necessary, empty
it as described in
;
.
11. Confirm that background counts are within limits. If the system has been idle
for fifteen minutes or more, a Background count should be run immediately
prior to running any calibration specimens.
12. Confirm that the Operator ID number is entered.
Calibration Materials
Calibrator Requirements
• For commercial calibrators, follow the directions given in the package insert.
Be certain to carefully read and follow directions given for warming and
mixing.
• Auto-Cal Method: The calibrator should be cycled for a minimum of 6 and a
maximum of 10 consecutive runs when calibrating in either Open or Closed
Mode.
• Enter Factor Method: The calibrator should be cycled for a minimum of 6
consecutive runs. Additional samples and/or repetitions of the specimens
may be used to achieve calibration accuracy beyond NCCLS
recommendations.
• The Calibrator sample should contain a minimum of 5.0 mL. If necessary,
aliquot a sufficient amount of sample into a single tube for Closed Mode
processing.
Fresh Whole Blood Requirements
The International Committee for Standardization in Hematology (ICSH) defines a
fresh blood sample as one available for processing less than four hours following
venous sampling.
Normal Sample
• Blood samples should be from the general patient population, with values for
all parameters which are within the laboratory’s normal range.
• All cellular morphology must be normal.
• No known interfering substances should be present (for example, lipemia,
icterus, medications).
• All samples must be properly collected in the EDTA anticoagulant used by
the laboratory.
• Each tube should contain at least 90% of the nominal collection volume of
blood.
• Samples should be at room temperature and mixed properly.
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...