10-94
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Troubleshooting and Diagnostics
Tables for Instrument Conditions and Messages
Section 10
The PC keyboard is not operational
– – – - Probable Cause – – – –
– – – - Corrective Action – – –
1. The computer is performing a
function that inhibits the keys.
1. No action required. Refer to the screen
for the current Status Box message.
2. There is an incomplete operator
entry.
2. Complete the operator entry or press
the Esc key on the PC keyboard.
3. A data transmission to the printer or
laboratory computer is in progress.
3. No action required. Refer to the
screen for the Status Box message.
4. Keyboard entry is not possible on
the displayed screen.
4. No action required. Refer to the screen
for the current Status Box message.
5. Computer, keyboard and/or circuitry
malfunction.
5. Power OFF the Analyzer. Then reset
the Keyboard cable on the rear panel
of the Data Module and power
Analyzer ON.
NOTE:
For the location of the
keyboard connector, refer
to
;
. If necessary,
contact your Abbott
Hematology Customer
Support Center.
The Sample Loader does not respond when the START LOADER,
RESUME, or RESET key is pressed
– – – - Probable Cause – – – –
– – – - Corrective Action – – –
1. The internal power cord connecting
the Sample Loader to the Analyzer
is loose or disconnected.
1. Contact your Abbott Hematology
Customer Support Center.
2. Circuitry malfunction.
2. Press
[FAULT REPORT]
in the main
DIAGNOSTICS MENU
and record any
fault listed.
3. Turn the instrument OFF and ON
again and prime the instrument.
4. If the problem persists, contact your
Abbott Hematology Customer
Support Center.
Table 10.4 Non-Functional Fault Conditions (Continued)
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...