CELL-DYN
®
3200 System Operator’s Manual
10-37
9140181H—October 2001
Section 10
Troubleshooting and Diagnostics
Loader Warning 145: unload area full
/BL
– – – – – – – – – – – – – – – – – Explanation/Action– – – – – – – – – – – – – – – –
The unload area contains 5 racks. NOTE: The Sample Loader halts.
No corrective action is necessary. When you wish to run additional samples in
the Sample Loader, remove the racks from the unload area and press the
[START
LOADER]
key.
Loader Warning 146: unload area nearly full
/BL
– – – – – – – – – – – – – – – – – Explanation/Action– – – – – – – – – – – – – – – –
The unload area contains 4 racks.
No corrective action is necessary. If a rack is being processed and there are no
other faults, the Sample Loader will continue to operate.
Loader Status 158: load zone empty
/BL
– – – – – – – – – – – – – – – – – Explanation/Action– – – – – – – – – – – – – – – –
The load zone empty sensor is not sensing the presence of a rack in the load zone.
NOTE: The Sample Loader halts.
When you wish to run samples in the Sample Loader, load tubes in racks and
press the
[START LOADER]
key.
Table 10.1 Status
Conditions (Continued)
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...