CELL-DYN
®
3200 System Operator’s Manual
4-11
9140181H—October 2001
Section 4
Performance Characteristics and Specifications
Hemogram Parameters
Precision specifications for the hemogram parameters are presented using
coefficient of variation (CV).
The stated CV% in the following Table represents the instrument precision at a
95% confidence level from N = 20 replicate runs based on the same sample.
WBC Differential Parameters
Precision specifications for the WBC Differential parameters are given as a 95%
confidence limit for N=20 replicate runs based on the same sample.
This confidence limit represents the allowable difference (plus or minus) of the
individual results from the mean of the 20 determinations (runs). Specimens
selected for the precision study must have means that fall within the specimen
range listed in both tables.
These samples should not display any of the following suspect Parameter Flags:
WBC or DFLT.
To determine the 95% confidence limit of a parameter:
1. Find the value listed in the Confidence Limit column for that parameter.
2. Add that value to the mean to determine the upper limit.
3. Subtract that value from the mean to determine the lower limit.
For example, if the mean of the Neutrophil parameter is 50, the 95% confidence
limit is 47.8–52.2.
Table 4.8
Precision of the Hemogram Parameters (N = 20)
Parameter
Specimen Range
CV
WBC (WOC)
4.1–11.0 x K/
µL
< 2.7%
WBC (NOC)
4.1–11.0 x K/
µL
< 2.7%
RBC
4.0–6.0 x M/
µL
< 1.5%
HGB
12.1–17.0 g/d
L
< 1.0%
MCV
81.0–100.0 fL
< 1.0%
RDW
12.0–16.0%
< 5.0%
PLT
150.0–400.0 x K/
µL
< 4.0%
MPV
5.0–10.0 fL
< 5.0%
RETIC %
0.5–2.7%
< 15.0%
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...