6-40
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Calibration Procedures
Enter Factor Method
Section 6
3. Transfer the Factor % Diff values from column 5 (calculated in Worksheet
2
)
to column 1 of Worksheet 3, Open Mode Calibration Range Criteria,
provided in
Manual Calibration Worksheets
at the end of this section.
NOTE:
Make copies of this worksheet as needed.
4. For each parameter, if the Factor % Diff is
equal to or less than
the value in
the Lower Limit column, then CALIBRATION IS NOT REQUIRED for that
parameter because the value is within range.
5. For each parameter, if the Factor % Diff falls
between
the upper and lower
calibration range, shown in the Calibration Range column, then
CALIBRATION IS REQUIRED.
6. For each parameter, if the Factor % Diff is
greater than
the value in the
Upper Limit column, there may be a computation error or an instrument
problem. In this case, do the following:
a. Recheck all numbers and calculations on Worksheets 1 and 2.
b. Determine if any component on the instrument was changed. This could
affect calibration. Such components include the Shear Valve, Optical
Flow Cell, Hemoglobin Flow Cell, or one of the syringes.
c. If a component was changed, then treat the result as if it fell within the
“calibration range” (even though it is greater than the upper limit).
CALIBRATION IS REQUIRED for that parameter.
d. If no component was changed and your calculations are correct, DO
NOT CALIBRATE. Confirm that all Pre-Calibration procedures were
completed and call the Abbott Hematology Customer Support Center for
assistance (at 1-800-CELL-DYN in the U.S.).
7. Based on the results from steps 1 through 6 above, do one of the following:
a. If any parameter needs to be calibrated in the Open Mode, go to
Entering
New Calibration Factors
—
Open Mode
.
b. If none of the parameters requires Open Mode calibration and Closed
Mode has not been calibrated, go to
Calibration Procedure — Mode to
Mode
to determine if Closed Mode needs calibrating.
c. If none of the parameters requires Open Mode calibration and Closed
Mode calibration has already been completed, press
[MAIN]
to return to
the
MAIN MENU
.
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...