Operating Instructions
Operating the Instrument
Activating The RUN Cycle
On the CS model, start the RUN cycle for either the Open or Closed Mode by
pressing the Touch Plate.
NOTE:
The door Assembly must be closed for the Closed Mode RUN cycle to be
activated.
On the SL model, Start the RUN cycle for Open Mode by pressing the Touch Plate.
Start the RUN cycle for Closed Mode by pressing the
[START LOADER]
key in the
RUN menu. For additional information, refer to
Sample Loader Description, Soft
Keys
in
.
Orderly Stops
The CS model, whether in Open or Closed Mode, automatically halts after each
sample has been processed unless the Touch Plate is pressed again to start a new
RUN cycle.
The SL model in Open Mode automatically halts after each sample has been
processed unless the Touch Plate is pressed again to start a new RUN cycle.
The SL model in Closed Mode automatically halts when no more racks are detected
on the load side and the last track has passed under the Tower to the Unload side.
Emergency Stops
To immediately halt all operation of the instrument, do one of the following:
1. Turn OFF the main power switch, located in the upper left corner of the back
panel.
2. Push the window section of the Tower Cover toward the instrument to release
the latches and break the safety interlock sensor connection. As soon as the
connection is broken, the instrument halts. It is not necessary to remove the
Tower Cover to stop operations. This method applies to both CS and SL
models. In addition, the CS model has a safety interlock sensor attached to
the swivel door mechanism which holds the sample tube.
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...