![Abbott CELL-DYN 3200 Скачать руководство пользователя страница 652](http://html1.mh-extra.com/html/abbott/cell-dyn-3200/cell-dyn-3200_system-operators-manual_450985652.webp)
14-64
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Reticulocyte Package
Routine Operation
Section 14
3. If the Specimen ID is found, the second
RETIC PATIENT SPECIMEN
screen is
displayed. All the available patient demographic information, and the RBC
value, are shown on the screen. Proceed to step 4.
If the Specimen ID is found but is more than 8 hours old, the third
RETIC
PATIENT SPECIMEN
screen is displayed and the operator may enter an RBC
value in this screen. Proceed to step 5.
If the Specimen ID is not found, the fourth
RETIC PATIENT SPECIMEN
screen
is displayed and the operator may enter an RBC value in this screen. Proceed
to step 5.
4. Press Y on the keyboard to save this demographic information.
NOTE:
If the Enter key was pressed in error, or if the patient demographic
information is incorrect, press the
[CANCEL]
key to return to the
RETIC SPECIMEN TYPE
screen, press the
[PATIENT SPECIMEN]
key,
return to step 2.
5. Press the Enter key on the keyboard to confirm and accept this information.
6. The
RETIC RUN RESULT
screen is now displayed. The area in the upper left-
hand corner displays the patient demographic information.
NOTE:
The patient demographic data can be added or edited on this screen
before the specimen is run, or from the
EDIT SPECIMEN
screen in
the Reticulocyte Data Log after the reticulocyte run is complete.
7. The prepared dilution(s) of the patient reticulocyte sample(s) can be run after
the control and background count results have met the laboratory’s criteria.
8. Open the well-mixed, prepared patient reticulocyte sample tube and immerse
the Open Sample Aspiration Probe in the sample.
9. Press the Touch Plate located behind the probe to start the cycle. The BUSY
indicator light on the Analyzer Status Indicator Panel will be illuminated in
yellow. The Status Box on the
RETIC RUN RESULT
screen will display
messages to indicate the various stages of the cycle.
10. Remove the sample tube when the beep sounds. The Wash Block moves
down the probe and cleans it.
11. When the rinse cycle is complete, the Wash Block moves up the probe. The
[NEXT RETIC]
key is now displayed, and the Touch Plate is no longer active.
The results of the reticulocyte run are displayed on the screen.
12. If automatic report printing has been specified, a report is printed according
to the options selected in the
SET UP MENU
. If automatic report printing has
not been specified, a report may be printed by pressing the
[PRINT REPORT]
key. Repeat this procedure for each subsequent reticulocyte sample.
NOTE:
To obtain a color printout, the color printing option on the
CUSTOMIZE PRINTED REPORT
screen must be turned ON.
13. Reticulocyte patient sample results can also be printed from the
RETIC
DISPLAY SPECIMEN
screen, available from the
RETIC DATA LOG
screen.
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...